SERVOTOUGH SpectraScan User Manual
02400001A/1
47
Figure 14 - Identification of RS-422 signals in instrument
This is shown in Illustration 4. Connect the cable to the individual push-down connectors as shown
in the figure. (NOTE: Although the board label says “RS-485”, this port will not function in multi-
drop bus mode. It will only work as a point-to-point link with a 4-wire RS-422 device.)
The RS-422 port parameters are as follows:
Baud rate = 57600
Data bits = 8
Stop bits = 1
Parity = None
Figure 15 - Location of RS-422 connector on system
board