5.21.8.
Get Window Row Stop Number
Description
Returns the stop row number for the ROI.
Setting Type
Operational
Command
WIN:ROW:STOP?
Parameters
none
Return Values
value
Range
7 to 255
Type
unsigned integer
Example
WIN:ROW:STOP?
-- query command
191 -- return value
5.21.9.
Set Window Rectangle
Description
Sets the global corner row and column numbers for the ROI.
The Total rows and columns must be an even value for each: Start
on an EVEN number and end on an ODD number, and conform to
minimum size requirements.
Setting Type
Global
Command
WIN:RECT
<X1> <X2> <Y1> <Y2>
Parameters
<X left> <X right> <Y top> <Y bottom>
Return Values
None
Range
0 to 319 for SU640HSX for corner row value
0 to 255 for SU640HSX for corner column value
Type
unsigned integers
Example
WIN:RECT 16 8 217 111 -- set window row and column points
5.21.10.
Get Window Rectangle
Description
Gets the global corner row and column numbers for the ROI.
Setting Type
Global
Command
WIN:RECT?
Parameters
none
Return Values
Range
0 to 319 for SU640HSX for corner row value
0 to 255 for SU640HSX for corner column value