SU320CSX
4110-0446, Rev. C
Page 75 of 96
© 2017 UTC
Aerospace Systems
Date Printed: 3-Jan-2017
This document does not contain any export controlled technical data.
3
Expose Block
Error
Invalid exposure and /
or frame rate timing
1. Increase readout time by modifying exposure
and frame periods.
2. Reduce trigger rate.
3. Reset firmware, reboot camera, or send a
trigger command that causes a trigger
parameter change (see section
Error!
Reference source not found.
).
4
Data RX PLL
Lock Error
Internal error
Power cycle camera
5..15
Unused
Unused
Unused
16
I2C0 Error
Internal Error
Power cycle camera
17
I2C1 Error
Internal Error
Power cycle camera
18
I2C2 Error
Internal Error
Power cycle camera
19
I2C3 Error
Internal Error
Power cycle camera
20
System
Temperature
Alarm
Camera Temperature
is out of acceptable
range
1.
Reduce camera case temperature
2.
Disable TEC.
3.
Turn off camera and let cool.
21
FPA
Temperature
Alarm
FPA Temperature is
out of acceptable
range.
1.
Disable TEC.
2.
Power off camera and let cool.
22
Camera
Alignment Error
Internal Error
Power Cycle Camera
23
Corr DL
Suspend
Correction
Download in
Progress
No action required
24
AGC High/Low
OPR Error
AGC High and Low
OPR settings are
invalid.
Correct OPR High/Low OPR settings.