SU320CSX
4110-0446, Rev. C
Page 72 of 96
© 2017 UTC
Aerospace Systems
Date Printed: 3-Jan-2017
This document does not contain any export controlled technical data.
Parameters
none
Return Values
value
Range
0 to 65535
Type
unsigned integer
Example
FPA:COLS? -- query command
640 -- return value
5.17.10.
Get Focal Plane Array Number of Rows
Description
Returns the number of rows of the focal plane array.
Setting Type
Global
Command
FPA:ROWS?
Parameters
none
Return Values
value
Range
0 to 65535
Type
unsigned integer
Example
FPA:ROWS? -- query command
512 -- return value
5.17.11.
Get Elapsed Time Meter
Description
Returns the total amount of time the camera has been powered on
since production. User is unable to reset.
Setting Type
Global
Command
ETM?
Parameters
none
Return Values
value (Days Hours:Minutes:Seconds)
Range
0 - 2
32
seconds, or about 136 yrs
Type
String
Example
ETM? -- query command
Days:8 04:03:02 -- return value
5.17.12.
Set Application Timer
Description
Sets the status of a timer which is stored in volatile memory.
Resolution is 0.1 second. Register is lost if power is interrupted.
Setting Type
Global