SU320CSX
4110-0446, Rev. C
Page 30 of 96
© 2017 UTC
Aerospace Systems
Date Printed: 3-Jan-2017
This document does not contain any export controlled technical data.
5.5.1.
Restore Factory Configuration
Description:
Restores the factory default settings. The User Configuration
memory space is erased. Then the Factory Configuration is copied
to the User Configuration memory space. Finally, the Current
Session Configuration is reloaded from the User Configuration.
Most modifications made by the user will be lost. However, offset
correction values overwritten with CORR:OFFSET:CAL will
persist. Appendix B (separate document) lists the Factory
Configuration global and operational parameter values for your
camera model.
Setting Type
Command
Parameters
Return Values
Example
N/A
CONFIG:RESET
None
None
CONFIG:RESET
5.5.2.
Save Global Configuration
Description:
Overwrites the User Configuration global settings with the Current
Session’s global settings.
Setting Type
N/A
Command
CONFIG:SAVE
Parameters
none
Return Values
none
Example
CONFIG:SAVE
5.5.3.
Load Operational Configuration
Description:
Loads the operational settings for the specified operational slot. See
the separate Appendix B document for a table of OPR settings and
corresponding integration and nominal FPA gain settings. An error
will occur if an opr_number outside of the available range is used.
Setting Type
N/A
Command
OPR opr_number
Parameters
opr_number
Return Values
none
Range
0 to N (limited by the number of operational settings that currently
exist, see the separate Appendix B document for table of OPR
settings).
Type
unsigned integer