12
REF 360-0040-01 (F)
SECTION TWO
PUMP OPERATION
2.1 Setting/Verifying Flow Rate
• Equipment Set-Up
1)
Use a GilAir-3 or GilAir-5 pump with a fully charged battery pack.
2)
Attach tubing to the pump.
3)
Connect the collection device to the tubing.
4)
Connect the tubing from the collection media to a Gilibrator 2.
5)
Setup and turn on the Gilibrator 2.
Figure 2.1
Field Calibration/Flow Verification Equipment Set-Up
2
5
1
4
3
WARNING
5
4
3
1
2
Battery
Fault
Flow
Possible static hazard. Do not
rub with dry cloth. Substitution
of components may impair
intrinsic safety. Use only
specified GilAir-3 battery packs.
See Labeling for intrinsic safety
approvals.
Personal Air Sampler
Gil ir3
A
A
4
3
2
1
Flow Adj.
On
Off