27
REF 360-0040-01 (F)
APPENDIX B
SPECIFICATIONS
General Specifications
Controls ......................................................... Power Switch, Flow Control (all models), plus:
MODE/HOLD button (Clock Model)
MODE/HOLD, PROG,
▲
, &
▼
buttons (Program Model)
Indicators ...................................................... Display (LCD), Battery (Green) LED, Fault (Red) LED
Display Ranges ............................................. 0–9999 (Timing), 0–9 (Error Percentage)
Display Messages .......................................... LAST, CAL, SHUT/OFF (Clock & Program Models)
E, PC, P1–P6, OLD (Program Model)
Dimensions (GilAir-3) .................................. 3.6” (H) x 3.9” (W) x 2.0” (D)
90 mm (H) x 100 mm (W) x 51 mm (D)
Dimensions (GilAir-5) .................................. 4.1” (H) x 3.9” (W) x 2.0” (D)
103 mm (H) x 100 mm (W) x 51 mm (D)
Weight (GilAir-3) .......................................... 21 oz. (595 g)
Weight (GilAir-5) .......................................... 22.5 oz. (638 g)
Power Supply
Battery Pack .................................................. GilAir-3: 4.8 volt, 1.8 amp hour
GilAir-5: 6 volt, 1.8 amp hour
Battery Type .................................................. Rechargeable Nickel-Cadmium or Nickel-Metal-Hydride
Battery Charge Time ..................................... 14–18 hours
Expected Battery Life † ................................. 300–500 charge/discharge cycles
Estimated Battery Life † ................................ 2.5 years (< 20 hours weekly use)
1.5–2.5 years (20–39 hours weekly use)
1–1.5 years (40–60 hours weekly use)
†
Inactivity for extended periods may shorten nickel-cadmium battery life.
Battery life estimates are based on proper battery maintenance.
Operating Specifications
Constant High Flow Range .......................... GilAir-3: 850–3000 cc/min
GilAir-5: 850–5000 cc/min
Flow Control ................................................. External Flow Adjust (
±
5% of set point)
Constant Flow (Low Flow) ........................... 5–500 cc/min to 25” H
2
0 (640 mm Hg)
Multi-Flow (Low Flow) ................................. 1–750 cc/min, to 18”+3” H
2
0 (460+80 mm Hg)
Operating Temperature Range .................... 32
°
to 104
°
F (0
°
to 40
°
C)
PRESSURE RANGE
Pump Flow
8 Hr Run (“H
2
O)
Min. “H2O before fault
LPM
GilAir-3 GilAir-5
GilAir-3 GilAir-5
0.85
20
20
25
35
1
25
25
30
35
2
–
26
–
35
2.5
15
–
20
–
3
8
23
10
30
4
–
15
–
15
5
–
5
–
5