14
When IR Link is connected to the Docking Station II, the Figure 5 will be
displayed. But if it is connected using USB Disk, the “Logs” tab will not indicate.
7.1.1 Docking Station Tab
✓
Serial Number
–
Serial number of Docking Station II (cannot be changed).
✓
Version
–
The current firmware version.
✓
Auto Power-Off
–
The time when unit automatically goes to Save Power Mode. Min/max
time can be set 1 ~ 3,600 seconds.
✓
Maximum Bump Time
–
The maximum and minimum time spending for Bump Test
process. If unit cannot respond correctly before time expires, the process will stop and
result will be fail. Can be adjust min/max of 20 ~ 120 seconds.
✓
Location
–
The location where test has done. It is beneficial when downloading the data
logs. Can add 12alphanumeric character.
✓
Gas Expiration Date
–
Informs the date of expiration of gas in installed cylinder. Must be
updated after installing a new one.
✓
Gas Bottle Lot #
- Displays lot number of installed cylinder. Must be updated after
installing a new cylinder.
✓
Info Read/Write
– By pressing the “Read” it displays current setting of the Docking
Station II and will discard any unsaved changes in the Tab.
✓
Calibrate
–
If the box is not selected, calibration process will not proceed by pressing
Calibration button. However, it will still update firmware and configuration of SGT.
✓
Bump Test
- If the box is not selected, Bump test process will not proceed by pressing
Bump Test button. However, it will still update firmware and configuration of SGT.
✓
Bump Result Decision
-
The dropdown list is selectable for “Beep”, “Concentration”, or
“Beep + Concentration”. The “Beep” option will run the bump test until the alarm
condition is met and the audible alarm is detected. The “Concentration” option will run
the bump test until the test gas concentration is verified. If the units are in Stealth Mode,
the Concentration option will be required for a successful bump test. The “Beep +
Concentration” option will run the bump test until the audible alarm is detected and the
test gas concentration is verified.
✓
Browse & Write
– To upgrade the latest firmware of Docking Station II, click “Browse”,
navigate to the firmware location and click “Open”. Click “Write” to begin upgrading
process.
Note:
Without clicking the “Write” button, configured and customized settings will not be
applied and neither be saved.
SGT Docking Station version II