3.2 Bump Test
Bump test is the process which checks and verifies the performance of the monitors and
ensures that sensors are correctly responding to the target gas.
To perform a Bump Test, follow the below:
✓
Insert the monitor(s) into the bay facing down.
✓
Slightly press down on the monitor until it locks into place.
✓
Press the Bump Test button for 2 seconds to execute Bump test.
✓
During the 20~30seconds, bay LEDs begin to light as the test progress.
✓
When test is finished, all (six) blue Bay LEDs illumination indicates if the SGT has passed
the test (see Figure 3).
✓
Otherwise, the third blue Bay LED illumination indicates if the SGT has failed the test (see
Figure 3).
SGT has passed the
test (six LEDs)
SGT has failed the test
(only 3
rd
LED)
Note:
Do not use the monitors that failed the calibration test.
4. Trouble Shooting
4.1 Bay Light does not turn on /test not performed
✓
Clean Monitor IR window.
✓
IR interference (bright lights/sunlight).
✓
Monitor in gas alarm state.
4.2 All tests fail
✓
Gas cylinder empty
–
check pressure gauge if it indicates at least 100psi/10bar.
✓
Gas Expired
–
check gas expiration date located on the cylinder.
✓
Wrong gas concentration(s)
–
gas cylinder and SGT concentrations not same.
✓
Wrong gas type
–
check the gas type.
SGT Docking Station version II
Figure 3