6.1 Connecting with SENKO IR Link

SENKO IR Link is a device which conveniently configures and communicates with SGT Docking 
Station II using the IR communication in Bay Unit 1.

To connect the IR Link, follow the below:

Plug the IR link into a USB port on PC.

Power on (green LED) the Docking Station II.

Place the IR Link facing up into the Bay Unit 1 (see Figure 4).

Double click and open the installed software.

Click “OPEN” button at the top right corner of the software interface.

The IR Link starts to flash rapidly.

After few seconds, “Read complete” message will be displayed, and current settings of 

Docking Station II will be showed on the interface.

Figure 4

Note: 

If “Send 

timeOut

” message displays, the IR Link has failed to communicate with 

Docking Station II. In this case, make sure IR Link screen is aimed at IR screen of Bay Unit 1 
and check Power of Docking Station II is on.

6.2 Connecting with USB

To connect the IR Link, follow the below:

Open the Cylinder compartment.

Find the USB port (see Figure 1) and remove the USB Disk.

Plug the USB into to PC.

Double click and open the installed software.

Select the “USB Disk” in the upper middle

-left corner.

Select the location of USB Disk and click “OK” button.

SGT Docking Station version II

Summary of Contents for SGT

Page 1: ...SGT Docking Station Version II User s Manual SGT Docking Station version II...

Page 2: ...ation of Gas Cylinder 3 Testing 3 1 Calibration 3 2 Bump Test 4 Trouble Shooting 5 Docking Station II Manager Software 5 1 Software Installation 6 Connecting to the SGT Docking Station II 6 1 Connecti...

Page 3: ...ice may cause a sudden reading change and cause the device malfunction Do not leave or use the device at the watery place or at where there is any liquid Do not use a device which has been failed the...

Page 4: ...arge fleets It also maintains all the important event logs that demonstrate users are working within company requirements 4 1 2 Instructions and Parts 1 1 Description SGT Docking Station version II a...

Page 5: ...x 21 3cm 20 66 x 17 20 x 8 40 Weight Without cylinder 11 0kg Temperature 41 to 104 F 5 to 40 C Warranty Full 2 years Battery Type Rechargeable Lithium Ion Battery Life Up to 1 000 Bump Tests LEDs 6 B...

Page 6: ...six blue LEDs see Figure 2 which indicates the process status Each of them will turn on by order from left to right while testing After testing is finished make sure all six bays are turned on otherw...

Page 7: ...ompartment Check the pressure gauge on the front panel Re install thumbscrews After installing a gas cylinder ensure USB is attached 3 Testing 3 1 Calibration Calibration is the comparison of measurem...

Page 8: ...has passed the test see Figure 3 Otherwise the third blue Bay LED illumination indicates if the SGT has failed the test see Figure 3 SGT has passed the test six LEDs SGT has failed the test only 3rd L...

Page 9: ...s The software can adjust and make changes to the Docking Station and the SGT units using SENKO IR Link and or partly USB Disk 5 1 Software Installation After downloading the file navigate to the save...

Page 10: ...10 The Installation will be ready to begin Click Install SGT Docking Station version II...

Page 11: ...s only when it is placed into the Bay 1 USB drive must be installed to store the test results Do not open the software if IR Link is not plugged to PC When the Docking Manager software opened the fiel...

Page 12: ...fter few seconds Read complete message will be displayed and current settings of Docking Station II will be showed on the interface Figure 4 Note If Send timeOut message displays the IR Link has faile...

Page 13: ...are Upgrade Self Test interval 0 20 hours Display mode Remain life Measuring mode Reading or Writing configuration on Docking station Enable Disable Cal Bump Test Serial Number Firmware Version Maximu...

Page 14: ...cking Station II and will discard any unsaved changes in the Tab Calibrate If the box is not selected calibration process will not proceed by pressing Calibration button However it will still update f...

Page 15: ...it can display remain lifetime of unit or measuring gas concentration Gas Concentration This allows a user to enter amend correct concentration of the installed gas cylinder on Docking Station II It i...

Page 16: ...fresh button to view and download the latest logs After clicking Refresh button all logs will be available to view see Figure 7 To download the logs follow the below Plug the IR Link into a USB port o...

Page 17: ...esting or calibrations Any unauthorized attempt to repair or modify the product or any other cause of damage beyond the range of the intended use including damage by fire lightening water damage or ot...

Reviews: