7.1.2 SGT Tab
✓
User ID
- The User ID can be used to add a usage message.
✓
Calibration Interval (Days)
–
The calibration reminder informs every fixed day (can adjust
0 (n/a) ~ 365).
✓
Bump Interval (Days)
–
The Bump test reminder informs every fixed day (can adjust 0
(n/a) ~ 365).
✓
Self-Test Interval (Days)
–
The Self-Test reminder informs every fixed hour (can adjust 0
(n/a) ~ 20).
✓
Low Alarm & High Alarm
–
The 1
st
and 2
nd
alarm set points for each gas.
✓
Display mode
–
Unit can display remain lifetime of unit or measuring gas concentration .
✓
Gas Concentration
–
This allows a user to enter/amend correct concentration of the
installed gas cylinder on Docking Station II. It is necessary to adjust for all tests.
✓
SGT Option Info Read/Write
– By pressing the “Read” it displays current setting of the
SGT and will discard any unsaved changes in the Tab. By clicking the “Write”,
all
configured and customized settings will be saved.
✓
Browse & Write
– To upgrade the latest firmware of SGT, click “Browse”, navigate to the
firmware location and click “Open”. Click “Write” to begin upgrading process.
SGT Docking Station version II