Operation • 79
17
-1
0-
05
-3
53
-0
02
g_
10
-2
02
0
B
Connecting the measuring mat
The measuring mat is intended for bioimpedance measurement by the 8-
point method (measurement of the whole body) with the patient in a lying
position. Measurement by the 4-point method (right half of the body) is like
-
wise possible.
1. Lay the measuring mat on the patient's knees.
►
The side with the writing on it should be facing upwards
►
The keypad should be facing the user
NOTE
If you wish to measure using the 4-point method, attach the electrode
cables on the keypad for the measuring mat to the hand and foot
electrodes of the right half of the body.
2. Connect the measuring mat to the electrodes.
►
Arrange the electrode cables so that they do not cross over one
another
►
Place the push-button adapter on the electrodes (black: Proximal,
red: Distal)
3. Ask the patient to do the following during the measurement:
►
Hold arms and legs away from the body
►
Lie still
4. Continue with the measurement as described in the relevant section:
►
Step 7. of the measurement with active WiFi direct connection
►
Step 6. of the measurement without active WiFi direct connection