9. FUNCTION CHECK
9.1 Device
►
Perform a function check prior to each use.
A complete function check includes:
• visual inspection for mechanical damage
• checking the alignment of the device
• visual and function check of the display elements
• function check of all the controls shown in the section entitled
“Overview”
• function check of optional accessories
If you notice any faults or deviations during the function check, first try to
resolve the error with the aid of the section entitled “Troubleshooting” in this
document.
CAUTION!
Personal injury
If you notice any faults or deviations during the function check which
cannot be resolved with the aid of the section entitled “Troubleshoot
-
ing” in this document, you may not use the device.
►
Have the device repaired by seca Service or by an authorized ser
-
vice partner.
►
Follow the section entitled “Servicing” in this document.
9.2 COVIDIEN
TM
GENIUS
®
3 in-ear thermometer
For the COVIDIEN
TM
GENIUS
®
3 in-ear thermometer, the manufacturer
CardinalHealth
TM
recommends a function check if one of the following
situations applies:
• In-ear thermometer was not used in accordance with purpose
• In-ear thermometer was dropped
• In-ear thermometer was stored below -25 °C or above +55 °C
10.SERVICING
10.1 Device
The measuring technology of this device must be checked every two years.
We recommend servicing the whole device as part of this check.
NOTICE!
Incorrect measurements as a result of poor servicing
►
Have servicing and repairs carried out exclusively by seca Service or
by an authorized service partner.
►
You can find service partners in your area at www.seca.com or by
sending an e-mail to [email protected].