Setting up peripherals • 177
17
-1
0-
05
-3
53
-0
02
g_
10
-2
02
0
B
Setting up automatic export
This function allows you to transmit measuring results to an EMR system
automatically. The export takes place as soon as data are generated or
amended on the device. To be able to use this function, an interface to the
seca analytics 115
PC software must be set up in your EMR system.
1. Establish which data format your EMR system accepts.
2. Press the
Peripherals
tab.
3. Press the
seca connect
button.
4. In the
Connect Mode
drop-down menu, select the option
seca analytics 115 TCP
or
seca analytics 115 UDP
.
5. Press the
button.
The
seca analytics 115 configuration
page is displayed.
6. Ensure that the
Automatic synchronisation
checkbox is activated.
7. Press the desired export format.
•
Automatic export (CSV)
•
Automatic export (EMR)
8. To adopt the settings, proceed as follows:
a) Press the
button
b) Press the
button
The settings are adopted.
9. Set up an interface to the
seca analytics 115
PC software in your EMR
system.