Clean the Assembly
CAUTION: Potential System Damage. Do not insert a wire or metal brush into the aperture
on the OptiFlow
TM
heater to avoid damaging the aperture.
1. Turn off the SelexION
®
technology, if it is installed.
2. Remove the nano cell curtain plate.
3. Loosen the two retaining pins that secure the nano cell heater assembly. Refer to
Figure 6-10 Nano Cell Heater Assembly Contacts and Retaining Pins
Description
Item
Retaining pins
1
Contact pins
2
Tightening sockets
3
4. Remove the nano cell heater assembly.
Note:
After the nano cell heater assembly and nano cell curtain plate are removed, make sure that they are
stored in the nano cell holder provided.