2. Use a PEEK tube cutter to cut a 2.5 cm piece of the 1/16 inch o.d. green FEP sleeve.
3. Insert the sleeve through the PEEK nut and the wide end of the ferrule, as shown in
, with the end
of the sleeve protruding 2 mm from the end of the ferrule.
Figure 4-4 Preparing the Sprayer Fitting
4. Insert the assembled fitting and the prepared sleeve into the rear of the sprayer head, and then tighten slightly.
Figure 4-5 The Sprayer Fitting Installed