4. Insert the high-voltage rail into the bracket and then push the rail forward until it stops.
Figure 4-7 Alignment Points on the Bracket
Note:
Resistance will be felt during installation of the rail. The resistance is caused by the O-ring, which
seals the Gas 1 connection.
5. Tighten the rail thumbscrew until it is snug. Do not overtighten it.
Connect the Sample Line
This procedure provides instructions for using fused silica tubing for the sample line. To use the PEEK-lined fused
silica tubing, which does not require cutting or sleeves, refer to
Connecting the Syringe Using PEEK-lined Fused
1. Cut a 3 cm length of the green PEEK sleeve.
2. Insert the syringe needle in the green PEEK sleeve.
3. Remove the stainless steel nut and ferrule from one end of the syringe union.
4. Insert the needle and sleeve into the stainless steel nut and ferrule, pushing the sleeve in as far as it will go,
and then insert the syringe needle.
5. Use two 1/4 inch wrenches to tighten the nut.
6. Insert one end of the fused silica tubing into the orange PEEK sleeve.
7. Cut the end of the fused silica tubing and then clean it with a wipe dampened with methanol or isopropanol.
8. Remove the stainless steel nut and ferrule from the other end of the syringe union.
9. Insert the sleeve and tubing into the stainless steel nut and ferrule until they protrude about 2 mm from the
tip of the fitting.
10. Insert the fused silica tubing, sleeve, nut, and ferrule, into the syringe union, pushing the sleeve in as far as it
will go.