Remove the Monitor
Remove the Monitor from the 4500, 5500, 6500, 6500
+
, and
TripleTOF
®
Systems
WARNING! Electrical Shock Hazard. Use only power supply model
HES10-12010-0-7 with the NanoSpray
®
III ion source.
1. Turn off the monitor.
2. Disconnect the monitor power supply cable from the power bar.
3. Disconnect the video and power cables from the monitor.
4. (Optional) To remove the bracket from the mass spectrometer, loosen the two 5 mm hex screws that hold the
bracket on the mass spectrometer, and then push the bracket up and pull it out to remove it.
5. Loosen the thumbscrew that holds the monitor on the bracket, and then remove the monitor from the bracket.
Figure 6-6 Bracket Assembly
Description
Item
Thumbscrew
1
Installation holes
2
Operator Guide
NanoSpray
®
Ion Source
RUO-IDV-05-0811-F