• For the NanoSpray
®
interface, type 225.
10. Wait 5 minutes for the interface heater to reach the correct temperature.
To determine whether the temperature has been reached, monitor the mass spectrometer detailed status by
double-clicking the mass spectrometer icon on the status bar. When the correct temperature is reached, the
Interface Heater Status
is
Ready
.
11. Let the interface bake at least 12 hours to eliminate any chemical contaminants.
Clean the Ion Source
WARNING! Electrical Shock Hazard. Remove the ion source from the mass
spectrometer before starting this procedure. Follow all electrical safe work
practices.
CAUTION: Potential System Damage. Be careful not to damage the emitter tip. It is fragile.
Clean the surfaces of the ion source after a spill or when they become dirty.
Tip!
When removing the high-voltage rail to reassemble the ion source head, inspect for residue under the rail.
Use a slightly damp, lint-free wipe to remove any residue from under the rail and from the outside of the O-ring.
Inspect the O-ring for damage. If it is damaged, replace it with a spare 1.5 mm by 1 mm O-ring (supplied).
Prerequisite Procedures
•
1. Remove the ion source from the instrument.
2. Wipe the surfaces of the ion source with a soft, damp, cloth.
Clean the Nano Cell Heater Assembly
Required Materials
Note:
U.S. customers can call 877-740-2129 for ordering information and inquiries. International customers
.
• Powder-free gloves (nitrile or neoprene recommended)
Operator Guide
NanoSpray
®
Ion Source
RUO-IDV-05-0811-F
62 / 92
Ion Source Maintenance