Kit
Part Number
Description
Item
Both
5015909
Tan PEEK sleeve (1/32 inch o.d., 0.015 inch
i.d.)
8
Consumables
1033299
Fused silica (100 cm, 75 µm i.d., 360 µm
o.d.)
9
Consumables
1003994
Orange PEEK sleeve (0.0625 inch o.d.)
10
Consumables
5015886
Syringe union
11
Consumables
1006549
Green PEEK sleeve (1/16 inch o.d., 0.030 inch
i.d.)
12
Both
1003988
100 µL syringe
13
Required Materials
• 2.5 mm hex screwdriver (PN 1034765)
• PEEK tube cutter (PN 011281)
• Fused silica cutter (PN 1006143)
• Two 1/4 inch wrenches (not supplied)
• HPLC-grade isopropanol or methanol (not supplied)
Tip!
For methods requiring a more robust syringe, an alternate syringe, PN 81075, is available from Hamilton
Company.
Remove the NanoSpray
®
III Head
WARNING! Electrical Shock Hazard. Remove the ion source from the mass
spectrometer before starting this procedure. Follow all electrical safe work
practices.
WARNING! Electrical Shock Hazard. Never operate the NanoSpray
®
ion source
without the illuminator, camera, stop, and covers properly installed. Never
touch the curtain plate or allow the emitter tip to contact the curtain plate.
If the mass spectrometer is operational and the ion source is installed, then
high voltage is present on the curtain plate, even if the X-Y-Z positioning
unit is moved away from the interface.
Operator Guide
NanoSpray
®
Ion Source
RUO-IDV-05-0811-F
26 / 92
NanoSpray
®
III Head Assembly and Installation