The following warnings apply to all of the maintenance procedures in this section.
WARNING! Hot Surface Hazard. Let the NanoSpray
®
ion source cool for at
least 60 minutes before starting any maintenance procedures. Surfaces of
the ion source become hot during operation.
WARNING! Fire and Toxic Chemical Hazard. Keep flammable liquids away
from flame and sparks and use them only in vented chemical fume hoods or
safety cabinets.
WARNING! Toxic Chemical Hazard. Wear personal protective equipment, including
a laboratory coat, gloves, and safety glasses, to avoid skin or eye exposure.
WARNING! Ionizing Radiation Hazard, Biohazard, or Toxic Chemical Hazard.
In the event of a chemical spill, review product Safety Data Sheets for specific
instructions. Use appropriate personal protective equipment and absorbent
wipes to contain the spill and dispose of it following local regulations.
WARNING! Electrical Shock Hazard. Avoid contact with the high voltages
applied to the ion source during operation. Put the system in Standby mode
before adjusting the sample tubing or other equipment near the ion source.
CAUTION: Potential System Damage. Do not lift or carry the ion source with one hand. The
ion source is designed to be lifted or carried using the molded grips on each side of the
ion source.
This section contains general maintenance procedures for the ion source. To determine how often to clean the ion
source or perform preventive maintenance, consider the following:
• Compounds tested
• Cleanliness of the samples and sample preparation techniques
• Amount of time an idle probe contains a sample
• Overall system run time
Operator Guide
NanoSpray
®
Ion Source
RUO-IDV-05-0811-F
44 / 92
6
Ion Source Maintenance