Operating and Maintenance Manual
Scientek Washer Model SW-6000
Revision-3, Sept. 2007
7-2
7.2 CYCLE T
EST
WARNING: FALL HAZARD
e.
To prevent falls, keep floors dry. Promptly clean up any spills or drippag
2.
Turn POWER switch to the ON position.
CAUT
Always
ssembly over quicklock con-
nec
n
not pos
ely wash
load.
ION:
position header shelf or shoe box rack a
tio in bottom of chamber before operation unit. If header or accessory is
itioned correctly, damage may result and unit will not effectiv
3.
Open chamber door and insert Quicklock cart (if supplied with machine) or ensure
Quicklock selector switch is in the OFF position. Close chamber door.
the machine for one or two complete cycles, verifying that they correspond to the
user's program and as outlined in
Section 2: Principles of Operation
.
.
During the cycle, verify the door switch operation as outlined in
Section 10.1.3.6.
7.
During the cycle, stop the washer immediately after the sump has completed filling with
er level in the sump. Adjust the fill delay timer
8.
efore
st the
the blower and
.
erature device and
compare it with the reading on the temperature display.
If it is out of adjustment, refer to Temperature Offset Adjustment Section 5.3.1.7.
4.
Press the START button. (Ensure that the machine door is closed and the
DOOR OPEN
indicator light is not illuminated).
5.
Operate
6
fresh water and check the wat
appropriately to compensate for different flow rates than those experienced during factory
tests and set-up.
During the cycle, stop the washer immediately after the end of the gravity drain and b
the flush drain commences. Check the level of water remaining in the sump and adju
power drain timer and gravity drain timer appropriately to compensate for different drain
back-pressures than those previously experienced.
9.
If the unit is supplied with a blower, devapormatic or drying package, start
orrect direction.
check that the blower is rotating in the c
7.3 TEMPERATURE CALIBRATION
Verify the temperature of the sump with a known accurate temp