Operating and Maintenance Manual
Scientek Washer Model SW-6000
Revision-3, Sept. 2007
5-16
5.3.1.6 Common Time Settings
The parameters are sorted in two list,
each having it “DOWN” and “NEW
VALUE” buttons.
The left list has the parameters for
adjusting the overfill (fill delay), drain
times, repeat dry, extra and emergency
exhaust time settings.
The right list is used to set up the
maximum time the system will wait
(hold) for temperature reaching the
guaranteed level, before generating an
alarm.
Use the respective ”DOWN” button to highlight the item you want to change and press the
“NEW VALUE” button. This will bring up a numeric keypad where you can enter a new
value.
ess “EXIT” button to go back to the Set-up menu screen.
This screen allows to set the duration for
of
the cycle, but will be continuously
pressed or the Emergency cable pulled.
The siren will be timed on any alarm and
ontinuous when the Emergency
cable is pulled.
The siren could be enabled/disabled by
touching the SIREN button.
(applies only if beacon & siren are
installed).
The wash solution re-circulation tanks’ operation parameters are adjusted on this screen: fill
delay from the tank, total re-fill capacity, maximum return duration, temperature set
point for
detergent tank, and number of times to recycle the wash solution before an automatic drain.
The drain duration for both tanks is adjusted on this screen.
Parameters on this screen are common to
all ten cycles.
Pr
the beacon and the siren.
The beacon will be timed at the start
activated if the Emergency button is
will be c