53
18.
Output Data
Appended Chart 1. Format of force response
Q±fffff±dddddddPLCSX
[Measuring value / Peak value]
m±fffff±dddddddPLCSYYMMDDhhmmss
[Saved data]
Description of respond data format
Q
Status of requested force data
f
Continuous output Measuring value (Approx. 2000data/sec.)
l
Continuous output Measuring value (Approx. 10data/sec.)
a Continuous peak value
h Continuous output -peak value
r Measuring value
p +peak value
n -peak value
1 1st peak value
2 2nd peak value
±fffff
4 digits force value with sign and decimal point
Ex., +123.4
+ddddddd
7 digits displacement value
with sign and no decimal point
Ex., +1234567
P
Unit number setting of force, 1 digit integer
0 to 5 (*)
L
Unit number setting of displacement,
1 digit integer
0 to 2 (*)
C
Comparator judgment
H Judgment: +NG
O Judgment: OK
L Judgment: -NG
E Overloaded
S
High / Low output
0 Less than No.1 / No.2
1 On and more than No.1
2 On and more than No.2
3 On and more than No.1 / No.2
X
Status of REC signal and mark point
0 No Rec input / No mark point input
1 No Rec input / Mark point input
2 Rec input / No mark point input
3 Rec input / Mark point input
4 Rec+Shift input / No mark point input
5 Rec+Shift input / Mark point input
YYMMDD
Saved date (YY
:
year /MM
:
month /DD
:
day)
hhmmss
Saved time (hh
:
hour /mm
:
minute /ss
:
second)
* Setting numbers and units are different depending on models. (Refer to page 50 of XFC command for detail.)