32
12.1.4. Saving single data
A single data is saved to USB memory when
‘
Single Data
’
is selected in the settings.
The data cannot be saved in the internal memory.
How to save
While
(MEM sign) is on, press
(SEND button) to save the data of which selected in
【
SEND
Functions
】
-
【
Send Data Select
】
. The message
‘
Data Saved
’
appears.
*See [18.3. File Format saved in then USB memory (HTGA only)] for the format of saving data.
*This operation creates a file in USB memory and adds data to it as repeated.
* In case the USB memory is disconnected and reconnected or the power turned off and back on, another
new file is created with this operation.
*It may pause a while before
‘
Data Saved
’
message appears when saving for the first time after USB
memory is automatically found.
12.1.5. Disconnect of USB memory
Please make sure to follow the direction below to disconnect USB memory from HTGA.
disappears when USB memory is ready to be disconnected.
Make sure to disconnect USB memory after
disappears.
●
Up to 100 data/sec. is saved in USB memory, while the sampling speed of HTGA is
2000 data/sec. The measuring value can differ between one displayed on HTGA and
one saved in USB memory because of the speed difference.
●Optional software Force Recorder is recommended for measurement with sudden force
change such as destruction test. Force Recorder can receive 2000 data/sec the same
speed of HTGA series.
●Do not disconnect USB memory during saving.
●Please make su
re to follow the direction to disconnect USB memory, otherwise data can
be lost.
●
The sign of
may not light after USB memory is repeatedly connected and
disconnected. Please turn off the power and turn it back on. An error message may
appear otherwise.
USB Disconnect
USB Memory
Program Menu