35
12. Output
12.4. Analog output
12.4.1. Analog output: D/A (standard spec.)
Analog voltage is always output depending on measuring force value. (+/- 2V when max. torque is
applied.)
Torque value can be recorded at real time by connecting to external equipments with analog cable
(excluded).
Analog output
Data update: 2000 data / sec.
Zero adjustment: /-20mV
Accuracy: 1% or less
*Connect to the external equipments with resistance 1k
Ω
and more.
●The analog output is unstable when the introduction message shows up on
the display. Please use the analog output during measurement.
12.4.2. Analog output: RAW (optional spec.)
The raw analog data is output without digital processing.
The response speed is fast, but zero reset is invalid. (Noise may also be detected as the data is not
filtered.)
Output voltage is approx. +/-1v when max. force is applied.
* Connect to the external equipments with input resistance 1k
Ω
and more.