5.1 Service Interval
5.1.1
Safety notes
The device must be serviced in regular intervals. The test results must be document-
ed and be compared with the values in the accompanying documents.
Maintenance work not described in this chapter, e.g. battery changes, may only be
accomplished by a qualified technician authorised by SCHILLER AG.
The following table gives information about interval and competence of maintenance
which can be required.
Interval
Service
Responsible
Every 6 months
• Keyboard test
• LED test
• Visual inspection of the unit and cables
"
User
Every 12 months
• Every servicing of the six month interval
• Functional tests according to the Service Handbook.
• Electrical safety tests according to IEC 60601-1, Clause 18 and 19
"
By SCHILLER AG author-
ised technician
Every 24 months
• Every servicing of the 6- and 12 months interval.
• Every measuring test and calibrations according to the service hand-
book.
"
By SCHILLER AG author-
ised technician
!
Danger of electrical shock. When working on a open device connect device al-
ways on a isolation transformer.
!
Follow the procedures for the prevention of accidents and environmental protec-
tion according your national guidelines.
Observe precautions for handling electrostatic sensitive devices when opening the
device.
Summary of Contents for AT-101
Page 1: ...A 0 ervi e andboo3 Art no 7 5400 Rev a 400 A 0 7 annel E G Devi e ...
Page 6: ...Page 4 Art no 2 540037 Rev a ...
Page 42: ...5 Care Maintenance 5 2 Functional test A 0 Page 40 Art no 2 540037 Rev a ...
Page 63: ...Page 61 Appendix 8 A 0 Service Handbook Checklist 8 1 Art no 2 540037 Rev a Other remarks ...
Page 64: ...8 Appendix 8 1 Checklist A 0 Page 62 Art no 2 540037 Rev a ...