Operation
3
A"#$0$
Service Handbook
ECG Settings
3.3
A
rt
.-
no
.:
2
.5
4
0
0
37
R
ev
.:
a
3.3.2
Automatic Format 1 and 2 External Printer
Two separate Auto formats can be defined for a external printer.
Press the SELECT softkey to choose from the following options:
Parameter
Options
Description
ECG Printout
No Printout
No printout of the ECG given at the end of an auto mode recording (the
recording can be stored in the memory and printed at a later time if re-
quired).
4x3 + 1 Rhythm
Leads are printed in a 4 x 3 format at 25mm/s, with the selected rhythm
lead at the bottom of the page at 25mm/s.
1x12 at 25 mm/s
Leads are printed in a 1 x 12 format at 25mm/s - no rhythm lead printed.
8x5 s + 4x10 s
The first 8 leads printed for 5 seconds and the last 4 leads printed for 10
seconds.
Short at 25 mm/s
Leads are printed in short form (1 sheet/lead group) at 25 mm/s.
Short at 50 mm/s
Leads are printed in short form (1 sheet/lead group) at 50 mm/s.
Long at 25 mm/s
Leads are printed in long form (2 sheets/lead group) at 25 mm/s.
Long at 50 mm/s
Leads are printed in long form (2 sheets/lead group) at 50 mm/s.
Average Cycles
No Printout
No printout of average cycles
4x3, 25mm/s + 2 Rhythm
Leads are averaged over the entire 10 second recording and printed in
4 groups of 3 leads at 25mm/s, with the two selected rhythm leads at the
bottom of the page at 25mm/s.
4x3, 50mm/s + 2 Rhythm
Leads are averaged over the entire 10 second recording and printed in
4 groups of 3 at 50mm/s, with the two selected rhythm leads at the bot-
tom of the page at 25mm/s.
2x6, 50mm/s + 2 Rhythm
Leads are averaged over the entire 10 second recording and printed in
2 groups of 6 at 50mm/s, with the two selected rhythm leads at the bot-
tom of the page at 25mm/s.
Rhythm Lead 1
I, II, III, aVR, aVl, aVF, V1 to V6 Select any lead.
Rhythm Lead 2
I, II, III, aVR, aVl, aVF, V1 to V6 Select any lead.
Measurements
On/Off
Select on or off to print a detailed table of measurement results
Markings
On/Off
Select on or off to print reference markings on the ECG average cycle
print. A vertical marker shows the beginning and end of P wave and
QRS, and the end of the T wave
Interpretation
On /Off
Select on or off to print interpretation statement
The sensitivity of the interpretation and the printing/not printing of abnormality state-
ments can additionally be set for the interpretation. These are described in paragraph
General
Lead
Interpretation
Filter
Auto. Format 2 external
Auto. Format 1 external
external
Auto Format 2 internal
Automatic Format 1
internal
ECG Settings
NEXT
Summary of Contents for AT-101
Page 1: ...A 0 ervi e andboo3 Art no 7 5400 Rev a 400 A 0 7 annel E G Devi e ...
Page 6: ...Page 4 Art no 2 540037 Rev a ...
Page 42: ...5 Care Maintenance 5 2 Functional test A 0 Page 40 Art no 2 540037 Rev a ...
Page 63: ...Page 61 Appendix 8 A 0 Service Handbook Checklist 8 1 Art no 2 540037 Rev a Other remarks ...
Page 64: ...8 Appendix 8 1 Checklist A 0 Page 62 Art no 2 540037 Rev a ...