4
System Settings
4.2
Unit
A"#$0$
Page 24
A
rt
.-
no
.:
2
.5
4
0
0
37
R
ev
.:
a
4.2 Unit
"
Press softkey UNIT
Parameter
Options
Description
User Identification (User ID)
Enter user ID
The user identification is printed on all recordings. The user ID can be
the department, doctor or hospital etc. Select User ID and a blinking cur-
sor is present - enter up to 30 characters via the keypad.
Note
If the unit is reset to the default settings (see following), the user identifi-
cation must be re-entered.
Date
dd-mm-yy
Enter the date in the format day.month.year. When set, the ENTER key
must be pressed to confirm.
Time
hrhr/minmin
Enter the time using the standard 24 hr notation. When set, the ENTER
key must be pressed to confirm.
Language
Deutsch
English (Default)
Francais
Svensk
American
Italiano
Espanol
Portuges
Dutch
Several languages are already programmed into the unit. Select the lan-
guage for the screen display and for the printout. The language will also
set the units used by the system.
The difference between American and English is as follows:
American
Standard English
measurements in inches
measurements in centimetres
temperature in Fahrenheit
temperature in degrees centigrade.
Mains filter setting - 60Hz
mains filter setting - 50Hz
date order mm-dd-yy
date order dd-mm-yy
Start-up Screen
Patient Data
Resting ECG
Here you can specify the first screen to be displayed when the unit is
switched on. Select between patient data screen (for entry of new pa-
tient) or data acquisition screen (ECG).
Paper
A4
Letter
The external printer can print on A4 or letter size paper. Set according to
the paper used.
TEST
AND INFO
EXIT
COMM
UNIT
Summary of Contents for AT-101
Page 1: ...A 0 ervi e andboo3 Art no 7 5400 Rev a 400 A 0 7 annel E G Devi e ...
Page 6: ...Page 4 Art no 2 540037 Rev a ...
Page 42: ...5 Care Maintenance 5 2 Functional test A 0 Page 40 Art no 2 540037 Rev a ...
Page 63: ...Page 61 Appendix 8 A 0 Service Handbook Checklist 8 1 Art no 2 540037 Rev a Other remarks ...
Page 64: ...8 Appendix 8 1 Checklist A 0 Page 62 Art no 2 540037 Rev a ...