4
System Settings
4.4
Test and Information
A"#$0$
Page 28
A
rt
.-
no
.:
2
.5
4
0
0
37
R
ev
.:
a
4.4.1
Print Setup
Printout for
Internal and external
4.4.2
Communications Test
"
To obtain a printout press the PRINT SETUP softkey.
A printout of the defined settings will be produced and gives the following information,
depending on the installed software:
SETUP
EXIT
COMM
TEST
BASE
INIT.
SOFTWARE
Print Setup Menu
Parameter
Settings
ECG Format (1 and 2)
Speed
Default speed setting
Internal Z external
Auto printout
Long (ooo), Short (o) or Suppressed (-)
MECG
Average cycles as defined in auto ECG recording setup
(e.g. 4 x 3 (25 mm/s) + 2)
Rhythm leads
Leads selected for R1, R2 resp.
Measurements
Print - Enabled (+) or Suppressed (-)
Marks
Print - Enabled (+) or Suppressed (-)
Interpretation
Print - Enabled (+) or Suppressed (-)
Leads
Sequence
Standard (S) or Cabrera (C)
Signals
Printout of signals - Sequential or Simultaneous
Auto Centering
Enabled (+) or Suppressed (-)
Lead Group
Rhythm, V9, V4r, V6r, DAJ, ON (+) or OFF (-) for each lead group
Filter
Baseline Filter
0.05, 0.15 or 0.30 Hz
Mains Filter
50, 60 Hz or OFF (-)
Myogram Filter
25 or 35 Hz, ON (+) or OFF (-)
SSB Filter
Smoothing Filter Enabled (+) or Suppressed (-)
SSF Filter
Baseline Stabiliser Enabled (+) or Suppressed (-)
Interpretation
Sensitivity
Low (-) or high (+) sensitivity
(only with option Interpre-
tation)
A30
Patient age is assumed to be < 30 (-) or >30 (+)
U
'Unconfirmed report' is written (+) or suppressed (-)
Abnormal
Normal / Abnormal printed Enabled (+) or Suppressed (-)
Thrombolysis
ON (+) or OFF (-)
Comm
Baud rate
115200, 57600, 38400, 28800, 14400 or 9600. This is followed by parity set-
ting (Y/N), bits and number of stop bits.
Mode
Line or Modem
When this is selected, test options are given for the RS-232 communication port. Use
this test if the RS-232 port is suspected of malfunction. A special test plug is used to
carry out the UART test. See page
SETUP
EXIT
COMM
TEST
BASE
INIT.
SOFTWARE
Summary of Contents for AT-101
Page 1: ...A 0 ervi e andboo3 Art no 7 5400 Rev a 400 A 0 7 annel E G Devi e ...
Page 6: ...Page 4 Art no 2 540037 Rev a ...
Page 42: ...5 Care Maintenance 5 2 Functional test A 0 Page 40 Art no 2 540037 Rev a ...
Page 63: ...Page 61 Appendix 8 A 0 Service Handbook Checklist 8 1 Art no 2 540037 Rev a Other remarks ...
Page 64: ...8 Appendix 8 1 Checklist A 0 Page 62 Art no 2 540037 Rev a ...