Unit Settings
8.1
ECG settings
AT-10 plus
Page 52
A
rt.
-n
o.:
2.5
4004
3 Re
v.
: a
Interpretation
Interpretation Screen
Edit/ Enter interpretation.
Write Unconfirmed Report Yes or No. ‘Unconfirmed Report‘ is added/not added to the interpretation state-
ments on the auto ECG printout (if applicable).
Write Abnormal ECG
Yes or No. ‘Abnormal ECG‘ is added/not added to the interpretation statements
on the auto ECG printout (if applicable).
Sensitivity
High or low sensitivity. Low sensitivity will suppress certain non-specific ECG
diagnoses; this may be advisable when carrying out ECGs for screening.
Thrombolysis (option)
On or Off. Thrombolysis is the breaking up of a blood clot. When Thrombolysis
option is off, the interpretation text “
possible infarct or other abnormality
” is
disabled.
Rhythm Rec
(not
available at time of
print)
Rhythm Lead R1, R2, R3
and r1
Defines the rhythm leads used for rhythm recording
Format
Defines the printout format (and by default the speed of the trace). Select
between:
• 90 sec/page - printout every 90 seconds
• 5 min./page - printout every 5 minutes
• 10 min./page - printout every 10 minutes
Amplitude
Defines the amplitude of the trace. Select between:
• Normal
• Low (half amplitude)
Print current page
For Rhythm recording see user guide.
Print last page
For the recording procedure see user guide.
Start / Stop
Starts / Stops Rhythm recording. For the recording procedure see user guide.
Late Potentials
(Not available at
time of print)
High Pass Filter
Select between 25Hz, 40Hz and 80Hz - late potential analysis RR is a future
option and not available for this release
No. of QRS
RR Variability
(Not
available at time of
print)
Window Type
RR variability is a future option and not available for this release
No of RR
FFt Type
Phase - space plots
Correction of RR
Lower bound of VLF
Lower bound of LF
Lower bound of HF
Upper bound of HF
Signal Acquisition Start /
Stop
Parameter
Options
Description