4.10 Software
4.10.1
Installing software options
All software options are embedded in the unit and can be upgraded with a code. To
upgrade an option(s), obtain the option code from SCHILLER, (you will need to state
the serial number of the unit
(1)
to get the code), and proceed as follows:
1. Click the ‘Program Upgrade’ icon
(2)
to display an entry field to the right of the
icon
(3).
2. Enter the option code (obtained from SCHILLER).
3. Press the
Enter
key and the program option is installed.
4. To ensure that the option is saved, press the menu key and enter the Software
ECG
NIBP
Spiro
HW-Config.
I/O + ports
AT-10 plus
V 1.04
Ctms
RI5.73
S/N :
173.000057
11.01.2005
Copyright (c) 2004 by
SCHILLER AG, Switzerland
Service
Character Colour
Test
Battery
Software
Program Upgrade
Program Update
1
1
2
3
4
The number to the right code entry
(4)
, indicates the number of unsuccessful attempts
to upgrade the program. If this numbers appraises or exceeds approximately 10, the
unit will freeze and cannot be upgraded. If this happens the unit must be returned to
SCHILLER for reprogramming.