Functional Checks
5
AT-10 plus
Service Handbook
General unit integrity check
5.2
A
rt.
-n
o.:
2.5
4004
3 Re
v.
: a
Continuous printout
8. Press the
Man Start
key to start a continuous printout. Check the printout
changes when the following keys are pressed:
Lead Group*
Sensitivity (Amplitude)
Speed
9. To stop the printout, press the
Stop
key.
If any function is suspected of malfunction, calibration is not correct, or the printout is
not correct for any reason etc., the full functional check must be carried out -see fol-
lowing.