Page 15
Introduction
2
AT-10 plus
Service Handbook
External connections
2.3
A
rt.
-n
o.:
2.5
4004
3 Re
v.
: a
2.3 External connections
2.3.1
Back Panel
(1) RS-232 connector for Treadmill.
(2) RS-232 connector for Ergometer.
(3) Potential equalisation stud. The potential equalisation stud is used to equalise
the ground potential of the unit to that of any nearby mains powered equipment.
Use the hospital or building common ground for all mains powered units.
(4) Mains connector and fuse box (fuses: 2 x T 160 mA / 250 V).
(5) VGA connector for external monitor. Note that before an external monitor can
be used, the VGA output must be enabled in system settings (see page 57).
(6) RJ-45 Ethernet LAN connector (Local Area Network).
(7) USB connector for an external printer.
(8) RS-232 for an external BP unit.
(9) SD card slot for (removable) data storage (64MB).
(10) USB connector.
(11) RJ-11 telephone connector (with optional internal modem).
V
All externally connected hardware must be approved by SCHILLER. Connection
of any hardware not approved by SCHILLER is at the owner‘s risk. The unit
guarantee may also be invalid.
1
2
4
5
6
7
8
9
10
11
3
V
If an external printer, monitor or ergo device is connected to the AT-10 plus, the
potential equalisation stud must be connected to common ground when the AT-
10 plus is working on battery power i.e.when the mains lead (with grounding lead)
is not connected to the unit (see page 17).
The following connectors are situated on the communications module.