Page 59
Unit Settings
8
AT-10 plus
Service Handbook
General and unit settings
8.3
A
rt.
-n
o.:
2.5
4004
3 Re
v.
: a
Modem
At the time
of print this function
was not fully
defined.
Phone Number
Enter the telephone number preceded by `T` or `P` (tone or pulse).
A comma `,` gives a one second pause in dialling - this may be necessary for
example, if an outside line is required.
Phone User Name
Phone Password
Server
Page
User Name
Password
Country ID
Peripherals
Blood Pressure unit
Select the way the blood pressure will be taken. Select between
Off
(the BP
entry screen is not displayed automatically during the test, however, the BP
screen can be displayed at any time during the test by pressing the NIPB key),
Manual Input
(the BP screen is displayed at the time interval defined in the
protocol - see user guide - the BP in taken by an external BP device and
entered during the test, or
Ergo 900
(an external, digitally controlled blood
pressure unit). With this unit BP is taken a defined intervals defined in the
protocol. The latest BP measurement is displayed on the monitor and all BP
measurements are stored with the exercise recording.
Bicycle
Select the type of bicycle connected - analogue or digital as follows:
• ERG900 / 911 / Ergosana (digital)
• SECA CT100 mod. 545 (digital)
• Analogue 1V / 100W
Note: the communication protocol for most digital bikes is standard. If you have
a bike not specified above, one of these two protocols should work.
Treadmill
Select the type of treadmill connected - analog or digital as follows:
• TM425 / TM400ES km/h
• TM425 / TM400ES mph
• RAM 770CE km/h
• RAM 770CE mph
• MTM-1500 km/h
• MTM-1500 mph
Format SDRAM Card
Click the START icon to format the card. This can be used when a new card is
inserted.
All data on the card is lost when an SD card is re-formulated.
Parameter
Options
Description