Service Screen
4.10 Software
AT-10 plus
Page 30
A
rt.
-n
o.:
2.5
4004
3 Re
v.
: a
4.10.2
Updating the unit software
The unit software is updated using a PC program called SWUP. Before updating the
unit software, the update program must first be downloaded from the SCHILLER
server to a PC.
The procedure to update the unit software is a follows:
1. Access the SCHILLER extranet (details from the service department) and install
the software on your pc (Install_AT10plus.exe).
2. Connect the RS-232 cable assembly between the PC and the RS-232 on the side
of the AT-10 plus.
3. In the service screen click the
Program Update
icon (to enter receive mode).
4. Start the AT10 plus SWUP program from the computer Start menu.
5. Click the config icon (
1
).
6. Define pc settings:
– Select the pc com port (
2
) to which the pc cable is connected.
– Select baudrate = auto.
7. Click the update AT-10plus icon
(3).
8. The bargraph shows the progress of the download. After downloading, the device
will restart automatically.
• The operating system of the PC must be Win NT4, Win 98, Win 2000 or Win XP.
• Do not power off during update.
•
RS-232 cable assembly.
Program Upgrade
Program Update
1
2
3
CANCEL
AT-10plus