Schambeck SFD GmbH
Refractive Index Detector RI2012
RI2012 V1.0 06-2013.EN / 28.10.2013
54
http://www.schambeck-sfd.com
Put the hex-wrench into one of the opened hole at the back side of optical system.
■
NOTE
There are two adjustment screws for the mirror. Keep in mind that you might have to turn
the other screw. Anyway you should turn the screws not more that 1/5 - 1/8 turns.
Adjust the screws until the optical balance reaches a value of 0 ± 10 %%.
Now the adjustment for the used solvent is finished.