Schambeck SFD GmbH
Refractive Index Detector RI2012
RI2012 V1.0 06-2013.EN / 28.10.2013
13
http://www.schambeck-sfd.com
9.8 T
HE
E
NTER
B
UTTON
Press the
ENTER
button to open the selected sub menu and to accept changes you
performed on a parameter.
9.9 T
HE
IN
PORT
This is the entrance for the mobile phase coming from the column. Connect the capillary
coming from the column to this port using screw connectors.
9.10 T
HE
OUT
PORT
This is the exit of the mobile phase. Connect you waste bottle to this port using the PTFE
tube.
■
NOTE
Use only capillaries with an inner diameter of 1 mm for the exit of the mobile phase. By this
blockage can be prevented. Make also sure that the pressure limits for the flow cell (6 bar,
0.6 MPa) and the valve (2 bar, 0.2 MPa) are not exceeded.