56
SCAIME Manual : NU-eNod4B-ETH-E-236708-D.docx
Single slope adjustment
Allows acquiring stable measurement with a load on the scale to set the first segment. Performing physical calibration in
this way do not allows multiple segments calibration. Use this calibration method only if you cannot acquire the
segment calibration immediately after the Zero calibration (e.g.: big hopper). Zero calibration must have been done
before.
The steps of this single slope adjustment operation are the followings :
1.
Define the number of points to realize the calibration and the load values for each,
2.
Press
,
3.
Place the Load 1 and press
,
4.
Wait for the end of data acquisition,
5.
Repeat steps 3 and 4 (if several points have been defined).
Remark
: If you press
,
the operation will stop and the values will not be saved.
The latest valid calibration values will be restored.
Theoretical calibration
An automatic scaling to migrate from the factory calibration to the user calibration.