Remove the screw and remove the pan support.
§
Replace the soiled in-use dust cover and ensure
that the adhesive points on the new cover properly
adhere to the scale's surface. The dust cover must
be firmly attached.
Important Note:
If the dust cover does not adhere to the scale,
weighing results may be distorted.
§
Replace and refasten the pan support.
§
Position the weighing pan.