9
Calibrate/adjust the scale following installation and after moving it to a new location.
Features:
Calibration can only be started when:
– The scale is unloaded
– The scale is tared
– The internal weighing signal is stable
If these prerequisites have not been met, an error message is displayed when you attempt
to start calibration.
The required calibration weight is displayed.
Standard calibration weights for selected models: see page 7 for instructions on removal.
You can select other weight values by pressing the
v
key.
You can interrupt calibration by pressing the
e
key twice.
Example: Determining weight value
Step
Press Key
Display
1. Turn on the scale
e
key
2. Zero/tare the scale
t
key
3. Start calibration; the defined calibration
q
key
weight is displayed without a unit
(in this example, 200 g)
4. If necessary, select a different
v
key
calibration weight
5. Confirm calibration weight value
h
key
and start calibration/adjustment
Once the zero point has been stored, the
calibration weight value flashes on the display.
6. Place the required calibration weight
on the weighing pan
The value stops flashing if the calibration
weight is within the defined tolerances.
When the calibration weight value has been
stored, the symbol
W
disappears and the
stability symbol
w
is displayed.
7. Remove the calibration weight
8. This concludes the calibration process
Calibration/Adjustment