Disturbance
Possible causes
Measures
Temperature control not accurately
Temperature of media is not
Ensure that the surface temperature
controlled accurately.
sensor is correctly positioned under the
bag. Check that BIOSTAT
®
CultiBag RM
600 is rocking. If the temperature is still
not controlled properly, recalibration of
temperature measurement might be
required. Contact your local Sartorius
Stedim Biotech representative.
Temperature appears to be incorrect
Actual temperature shown in main menu
Make sure that surface temperature
does not correspond with independent
sensor is correctly po¬sitioned under
measurement or ambient temperature.
the bag. Check that BIOSTAT
®
CultiBag
RM 600 is rocking. If actual tempera-
ture is still not correct, recalibration of
temperature measurement might be
required. Contact your local Sartorius
Stedim Biotech representative.
Supplement
33