3.3.4 Room of Installation
1. The operator needs only access at the front of the BIOSTAT
®
CultiBag RM 600.
For calibration and maintenance, access from the rear side and left side is also
required.
The BIOSTAT
®
CultiBag RM 600 is mobile, but must not be moved with a filled
bag inside.
2. Place the BIOSTAT
®
CultiBag RM 600 on a horizontal, safe and stable floor.
3. Do not place any equipment close to the rocking table to avoid any blockage or
damage. Also do not place the BIOSTAT
®
CultiBag RM 600 to jut out from parts
of wall which could be a hazard risk.
4. The measurement of pH and DO is based on chemical sensors. The chemical pH
sensor is susceptible to photo bleaching, keep it covered. Avoid intensive light
such as sun beam shining on the BIOSTAT
®
CultiBag RM 600 with mounted
CultiBags.
Do not expose the Cultibag RM on direct sun beam.
3.4
Equipping the system
3.4.1 Mount BIOSTAT
®
RM Control Tower
1. Connect the BIOSTAT
®
RM Control Tower with BIOSTAT
®
CultiBag RM 600 using
included RS232 connection cable and power supply.
Do not modify the BIOSTAT
®
RM System and Cultibag RM 600 in any way.
Modification may affect the safety of device.
3.4.2 Surface temperature sensor, plug in
The surface temperature sensor is mounted at rocking table.
Connection cables have to be plugged in.
1. Correct placement means the surface of the sensor is visible on top so that it
is in direct contact with the CultiBag RM 600 film as shown in the picture.
2. Guide connection ribbon cables with plug through the opening of rocking
table at the front side.
Incorrect placement of the temperature sensor can cause overheating of the heater
plate and the bimetal will disconnect the thermal energy after reaching 50 – 60°C.