User interface
Liquid density
in m³
Density of medium at 1 bar (abs) and boiling temperature. To improve the accuracy of the in
-
dicated filling level, enter the liquid density at operating conditions. In this case, the operat
-
ing pressure density must be entered and changed manually (see [Operating pressure]). This
makes the filling level reading more precise.
Standard gas density
in m³
Gas density at 1 bar (abs) and 15 °C. The current values for liquid density, gas density in the
tank, and gas density in the low-pressure pipe are automatically converted and displayed.
Gas density in tank
and
Gas density in low-pressure pipe
in m³
−
Details on gas column correction
Gas density in the tank depends on the nominal pressure and assumed temperature in the
tank. Gas density in the low-pressure pipe depends on the nominal pressure and assumed
temperature in the pipe.
Note:
The gas density in the tank and in the low-pressure pipe are required for gas column
correction of the content reading. If no correction is required, set both densities to 0.
Mixture density
in m³
Mixture density refers to the density when filling the tank. The operating filling limit is derived
from this. It is automatically calculated on entering the liquid density.