Salvislab Thermocenter TC40/100 User Manual
- 11
Copyright© SalvisLab / Edition 30062016_2.0
5.8
Multi user levels
This new feature allows more than one user on a single device by separating their accounts, application data
and access rights.
User*
It is the lowest user level. This user level allows to run the unit, define custom
programs, set date and time, temperature measurement unit, restart procedure,
process data sampling time, check for unit errors and access to the basic informations
of the unit.
Calibrator*
The Calibrator level includes the user level capabilities and integrates the unit
calibration.
Administrator**
The Administrator level includes the User level capabilities. In addition it allows to
perform tests on the critical components of the unit, update software, access to the
system log, reset the user’s database and run a sterilization program. It also
guarantees the ability to modify some key working parameters of the unit such as
upper and lower deviation, DHCP, max adjustable temperature and buzzer delay. It
provides User administration rights.
Service***
It is the highest level. This access is granted only to qualified factory technicians for
service and upgrades.
5.9
Data connection
•
USB 2.0 Port
•
LAN-Port, RJ45 10/100 Mbit - optional
5.10
Door alarm
•
The door is monitored by a mechanical switch
•
It is not possible to start a manual process when the door is open; the message “door open” will appear
on the display
•
The heating process is interrupted when the door is opened during a manual process (heating off), as
soon the door is closed the heating process will start again (heating on)
5.11
Safety / Power Cut-Off
•
Independent temperature safety device class 3.3 according to DIN 12880:
+
Integrated digital under- and over-temperature protection which automatically follows the
set point value at a preset tolerance range, and switches the heating off in case of over- or under
temperature deviations
+
A built in safety controller takes over if the first protection is in state of error and shuts the oven
down in case of over-temperature
+
A mechanical over-temperature device provides additional safety
•
After power cut-off controller restarts automatically and last stored values remain in memory
•
High quality accurate PT100 temperature probes
•
Integrated door switch interrupts heater when opening the door