Salvislab Thermocenter TC40/100 User Manual
- 9
Copyright© SalvisLab / Edition 30062016_2.0
5
Introduction
5.1
Overview
•
The THERMOCENTER TC40/100 is an oven with a patented design
•
Robust “SwissTEQ” quality design
•
Thermocenter ovens ensure effective drying without damaging the material being dried
•
All functional elements are integrated in the door
•
High grade stainless steel chamber
•
Standard unit provided with one aluminium shelf
•
Chamber has rounded edges which allows easy cleaning
•
Minimized heat losses due to low thermal conductivity insulation
•
Exterior textured powder coated steel
•
SalvisTEQ controller with dynamic regulation and improved efficiency
•
Forced Air with an intelligent control of fan speed in a range between 60 and 100%
•
Capacitive touch display, intuitive and user-friendly
•
Multiple user levels
•
Drag&drop user lock function
5.2
Applications
•
The Thermocenter line is designed for all purposes of drying in a variety of laboratory fields
•
Controller with capacitive touch display and unlimited programming capabilities
•
A must when precise temperature distribution and a high accuracy are needed
•
Temperature range up to + 200°C/392°F
•
It can be used for research & development, quality control as well as for a wide range of industrial and
laboratory applications
•
Examples of usage: color fastness test for textiles, ageing test for plastics and foils, quality control of
electronic circuits, food analysis, dry sterilization in hospitals
5.3
Construction
•
Due to extremely compact all-in-door construction the Thermocenter saves valuable space in lab
•
Electro-polished stainless steel inner chamber, chemical resistant and highly durable
•
All edges rounded for easy cleaning and maintenance
•
Maximized shelves area (in comparison to inner volume): up to 8 shelves
•
Door seal easily removable for cleaning or replacing
•
Door handle fully integrated in the door
•
No hot surfaces
5.4
Controller
•
SalvisTEQ controller with dynamic regulation
•
The regulation unit controls the heating with a predictive regulation
•
Multi-point temperature calibration (up to 10 points)
•
Self identification of heating gradient
•
Real time clock with process times
•
Calendar-settable programs start (date/time)
•
Hold time (process-end shown as date/time): 0 - 999h 59m
•
Enhanced program mode (up to 99 programs and 99 steps per program)
•
Scheduler
•
Programs remain stored in memory even without external power
•
Import/export of programs over USB - optional: LAN
•
Real time visualization of current program´s process data
•
Export process data over USB - optional: LAN
•
Preselect restart procedure in case of power failure
•
Interal DU memory expandable via micro SD card up to 32GB - optional
•
Log of process data for 300.000 entries (approx. 400 hours at a sampling rate of 5s)