Salvislab Thermocenter TC40/100 User Manual
- 48
Copyright© SalvisLab / Edition 30062016_2.0
12.6.17.3
Alarm contact
The potential free contact is used for the alarm contact. This contact is closed in an error-free condition. If
an error appears, the contact is opened. The contact will stay open as long as the user acknowledges all
errors in the list.
i
This feature is only available with the corresponding option.
12.6.18
DAC
The device has two analogue outputs. They can be mapped to the system sensors.
The following options are possible:
Sensor
Formula
Temperature internal
Voltage [V] = temperature [°C] * 30 +1V
Temperature external
Voltage [V] = temperature [°C] * 30 +1V
Temperature ambient
Voltage [V] = temperature [°C] * 30 +1V
Fan speed
Voltage [V] = fan speed [%] / 10
12.7
Logs
Users have the option to export or delete logged process data.
1
Back
2
List of process logs
3
Delete selected process data
4
Export selected process data
5
Back
i
Only available with the corresponding option.
5
1
2
3