5-39
© Sage Instruments 2000
9100-2930-02 01/31/00
Sage Instruments 930i Communications Test Set
The display reads:
LO=0 MD=0 HI=0 00:00 START
K1
K2
K3
K4
2. Press K4 under
START
when you are ready to start testing. The display reads:
LO=0 MD=0 HI=0 00:00 RANGING
K1
K2
K3
K4
This display is only momentary and is replaced by the measurement display:
LO=0 MD=0 HI=0 00:01 STOP?
K1
K2
K3
K4
The
LO
value is the threshold setting, the
MED
setting is thr spread, and
HI
is thr2X
spread.
If you need to stop the test, press K4 under
STOP
. Otherwise the test will accumulate impulse noise as
selected under the
TEST LENGTH
display.
If the test is to be performed with holding tone over a loopback circuit, the 930i should be set to send
1020 Hz
at
-13.0 dBm
using the Send Tone function. If the received holding tone drops below
-40.0 dBm
during the test, the words
NO TONE!
appear above K4.
3. Press any soft key (K1-K4) to display the Phase Hits, Gain Hits, and Dropouts. The display
reads:
PH=0 GH=0 DO=0 00:01 STOP?
K1
K2
K3
K4
4. Press the Option Menu function key to exit.
Option Menu #:
11
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for 930i
Page 11: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 19: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 43: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 203: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 235: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...