© Sage Instruments 2000
9100-2930-02 01/31/00
5-120
Section V Option Menu Numbers
6. Press K2 under
FREQ
. The display reads:
FREQ:
2100 Hz DEFAULT EXIT
K1
K2
K3
K4
7. Use the numeric keypad to enter a new frequency, and then press K4 to exit to the
MODIFY
display.
• A valid frequency is between
20 Hz
and
5000 Hz
. The default is
2100 Hz
.
8. Press K3 under
TIMING
to change the timing and phasing. The display reads:
TIMING: DURATION PHASING EXIT
K1
K2
K3
K4
9. Press K2 under
DURATION.
The display reads:
DURATION: 2.2 SEC DEFAULT EXIT
K1
K2
K3
K4
10. Use the numeric keypad to enter a new duration and then press K4 to exit to the previous
display.
• A valid duration is between 0 SEC and 99.9 SEC. The default is 2.2 SEC.
11. Press K3 under
PHASING
. The display reads:
PHASE SHFT: ON 450MSEC DEFAULT EXIT
K1
K2
K3
K4
12. Press K1 under
ON
to toggle the phase shift on and off.
13. Use the numeric keypad to enter a new phase shift duration displayed above K2.
• A valid phase shift duration is between
10 MSEC
and
9999 MSEC
. The default is
450 MSEC
.
14. Press K3 under
DEFAULT
to set the phase shift duration to its default setting.
15. Press K4 to exit to the main display.
Option Menu #:
59
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for 930i
Page 11: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 19: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 43: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 203: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 235: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...