5-19
© Sage Instruments 2000
9100-2930-02 01/31/00
Sage Instruments 930i Communications Test Set
Option Menu #:
4
4. Press K2 to add a one-second pause, K3 to add a dial delay, OR K4 to select a testline. (Testline
allows you to select
100, 102, 105,
or
LOOPBACK
in PCM mode.)
5. Press the Option Menu function key to return to the main
SET RECEIVE
display, OR press K4
under
EXIT
to back out another step to the main selection display.
5-4.2 Receive
To activate the digit receiver:
1. Press K2 under
RECEIVE
from the
SET-UP
display.
for example:
REC #1 HOLD GATED
K1
K2
K3
K4
When the trunk is seized and digits are sent, the digits appear on the left side of the display.
2. Press K2 under
REC #1
to scroll through the available records.
3. Press K3 to toggle
HOLD
on/off. In the on position, the captured digits are held until K3 is pressed
a second time. The factory default is on.
4. Press K4 under
GATED
to select a trigger. The display reads:
TRIGGER: OPEN GATED SCAN
K1
K2
K3
K4
5. Press K2 to activate
OPEN
mode, K3 to activate
GATED
mode, or K4 under
SCAN
(when in PCM
mode only) to scan all PCM channels.
OPEN
receives digits regardless of the supervision state.
GATED
waits for an on–to–off-hook transition before the digit receiver is activated.
SCAN
scans all PCM channels until it sees one channel go off-hook and collects digits from that
channel. The 930i continues to monitor the active channel until it goes back on-hook or
until the requested digit records are full. It automatically scans the next call until K4 is
pressed again, discontinuing the function.
In GATED mode, the digit receiver waits indefinitely unless the correct signalling is received before
digits are sent. If digits can be heard with the speaker volume turned up but do not appear immediately
on the display, check your trunk type settings and sequence. Or use OPEN mode.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for 930i
Page 11: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 19: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 43: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 203: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 235: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...